Acceleration transducer mod. TA18S
FUNCTIONING
The TA18S transducer picks up seismically the absolute vibrations of the machine by being fitted directly to
the supports of the vibrating part; it supplies an output signal directly proportional to the vibration of the point
to which it is fastened. Such signal should subsequently be processed by one of the measuring channel of a
CEMB serial “T” or “N” processing unit.
TECHNICAL CHARACTERISTICS
Type of measurement : seismic (absolute vibrations)
Dynamic range : ± 50 g
Frequency response : ± 3 dB 0,5 ÷ 15000 Hz
± 10 % 0,8 ÷ 6000 Hz
± 5 % 1 ÷ 4000 Hz
Direction of vibrations : any
Sensitivity : 100 mV/g
Tranverse sensitivity : < 5 %
Reasonance frequency : 25 kHz
Protection against shocks : 5000 g pk
Power supply : 2÷20 mA - 18÷28 Vdc
Outlet impedance : < 150 ohm
Temperature range : -54°C ÷ +121°C
External casing material : stainless steel AISI 316-L
Mounting screws : standard = ¼”-28UNF-2B or
to be specified along with the order among those shown on dwg 58608-P
Weight : 88 g
Protection against external : IP65 EN60529/10.91
Connection : 2 pin connector MIL-C-5015 serie 3106/10, supplied as standard
ASSEMBLY
Make a threaded hole on the support to be tested. The surface where the transducer rests must be perfectly
smooth and flat. Is is advised to provide a film of silicon between the resting surface and the sensor.
NOTE : avoid hitting violently the transducer housing, the closure couple must be 2,7÷6,8 Nm.
MAINTENANCE
Any.
DIMENSIONS, FIXTURE AND CONNECTIONS
As per enclosed dwg. n° 58608-P.
Vue de la page 0